文化大學機構典藏 CCUR:Item 987654321/29147
English  |  正體中文  |  简体中文  |  全文笔数/总笔数 : 46833/50693 (92%)
造访人次 : 11846349      在线人数 : 555
RC Version 6.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
搜寻范围 查询小技巧:
  • 您可在西文检索词汇前后加上"双引号",以获取较精准的检索结果
  • 若欲以作者姓名搜寻,建议至进阶搜寻限定作者字段,可获得较完整数据
  • 进阶搜寻
    主页登入上传说明关于CCUR管理 到手机版


    jsp.display-item.identifier=請使用永久網址來引用或連結此文件: https://irlib.pccu.edu.tw/handle/987654321/29147


    题名: Validity of Standing Posture Eight-electrode Bioelectrical Impedance to Estimate Body Composition in Taiwanese Elderly
    作者: Lee, Ling-Chun
    Hsieh, Kuen-Chang
    Wu, Chun-Shien
    Chen, Yu-Jen
    Chiang, Jasson
    Chen, Yu-Yawn
    贡献者: 運教所
    关键词: body composition
    dual-energy X-ray absorptiometry
    elderly
    日期: 2014-09
    上传时间: 2015-01-20 13:52:57 (UTC+8)
    摘要: Background: Standing eight-electrode bioelectrical impedance analysis, which can be used to estimate percentage body fat (BF%) and lean soft tissue (LST) in the whole body and different body segments of elderly adults, is potentially an ideal method for clinical assessment of body composition.

    Methods: In this study, dual-energy X-ray absorptiometry (DXA) was used as a criterion method to validate a standing eight-electrode bioelectrical impedance analysis device BC-418 (hereafter abbreviated as BIA(8); Tanita Corp., Tokyo, Japan). LST and BF% were measured in the whole body and various body segments (upper limbs, lower limbs, and trunk) of each participant using BIA(8) and DXA; correlation and differences between the LST and BF% results measured in the whole body and various body segments were compared. A total of 77 individuals, 42 males and 35 females, aged 55.2-76.8 years, were included in the analysis.

    Results: The impedance indexes (h(2)/Z) of hand to foot, and upper and lower limbs of the left side of the body measured by BIA(8) were highly correlated with the LST values in the corresponding body segment measured by DXA (r = 0.96, r = 0.92, and r = 0.88, respectively; all p < 0.001). LST values of the whole body and various body segments of participants measured using BIA(8) were highly and significantly correlated with the corresponding DXA data (all r > 0.88, p < 0.005); the whole body and segmental BF% measured by BIA(8) and DXA also showed a significant correlation (r > 0.84, p < 0.005). In addition, the agreement between the results of BIA(8) and DXA was assessed by Bland-Altman analysis; the bias and SD were, respectively, 1.89 kg and -4.25% in limb LST, and 2.18 kg and 4.06% in whole body BF%.

    Conclusion: The results of this study showed that the impedance index and LST in the whole body, upper limbs, and lower limbs derived from DXA findings were highly correlated. The LST and BF% estimated by BIA(8) in whole body and various body segments were highly correlated with the corresponding DXA results; however, BC-418 overestimates the participants' appendicular LST and underestimates whole body BF%. Therefore, caution is needed when interpreting the results of appendicular LST and whole body BF% estimated for elderly adults. Copyright (C) 2014, Taiwan Society of Geriatric Emergency & Critical Care Medicine. Published by Elsevier Taiwan LLC. All rights reserved.
    關聯: INTERNATIONAL JOURNAL OF GERONTOLOGY 卷: 8 期: 3 頁碼: 137-142
    显示于类别:[運動教練研究所] 期刊論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    index.html0KbHTML416检视/开启


    在CCUR中所有的数据项都受到原著作权保护.


    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回馈