    題名: Adjustable trapping position for single cells using voltage phase-controlled method
    作者: Wang, CC (Wang, Chun-Chih)
    Lan, KC (Lan, Kung-Chieh)
    Chen, MK (Chen, Ming-Kun)
    Wang, MH (Wang, Min-Haw)
    Jang, LS (Jang, Ling-Sheng)
    貢獻者: Dept Elect Engn
    關鍵詞: Cell manipulation
    Cell impedance measurement
    Voltage phase-controlled method (VPC)
    Dielectrophoresis (DEP)
    Adjustable trapping position
    日期: 2013-11-15
    上傳時間: 2014-02-21 14:57:50 (UTC+8)
    摘要: We present an advanced technique improving upon the micron-sized particle trap integrated in biochip systems using a planar structure to generate an adjustable trapping position by utilizing voltage phase-controlled (VPC) method and negative dielectrophoresis (nDEP) theory in high conductivity physiological media. The designed planar and split structure is composed of independent components of measuring and trapping micro-electrodes. Through different voltage configurations on the device, the trapped position of single particles/cells was selected and adjusted in vertical and horizontal directions. The numerical simulations verify our theoretical predictions of the effects at the various voltages. It shows that the trapped position can be adjusted in the vertical (0 to 26 pm) and horizontal (0 to 74 mu m) directions. In experiments, the single particles/cells is captured, measured, and then released, with the same process being repeated twice to demonstrate the precision of the positioning. The measurement results determined that particles at various heights result in different magnitude values, while the impedance error is less than 5% for the proposed electrode layout. Finally, the experiments are performed to verify that a particle/cell can be precisely trapped on the selected site in both the vertical and horizontal directions. (C) 2013 Elsevier B.V. All rights reserved.
    關聯: BIOSENSORS & BIOELECTRONICS Volume: 49 Pages: 297-304
    顯示於類別:[電機工程系] 期刊論文